Editorial Technical Reference

Precision balance

Laboratory weighing · Accuracy classes, verification interval, permissible errors, repeatability, eccentricity and minimum weight

A precision balance weighs laboratory and production samples to fractions of a gram. Ask three things first: the capacity and readability, the accuracy class and verification interval if it is used for legal or regulated weighing, and the repeatability and minimum weight that decide the smallest sample you can weigh accurately.

Electronic laboratory precision and analytical balances, from about 0.1 g down to 0.1 mg readability. Industrial platform and truck scales, retail scales, moisture analyzers and density kits are outside this page.

CNFX editorial research · Sources checked October 12, 2026 · 4 sources

Gloved hand placing a stainless check weight with tweezers on the pan of a balance inside a glass draught shield, with weight sets on a granite bench
Balance with draught shield and check weights

01

Check the accuracy class and permissible errors

OIML Recommendation R 76-1 classifies non-automatic weighing instruments as special (I), high (II), medium (III) or ordinary (IIII) accuracy by the verification scale interval e and the number of intervals n = Max/e. Class I needs e of at least 0.001 g and at least 50,000 intervals, with a minimum capacity of 100 e; class II with 0.001 g ≤ e ≤ 0.05 g needs 100 to 100,000 intervals and a minimum capacity of 20 e. The note adds that testing below e = 1 mg is not feasible in normal practice because of test-load uncertainty. [1]

Maximum permissible errors on initial verification are ±0.5 e, ±1.0 e or ±1.5 e depending on load: for class I, ±0.5 e up to 50,000 e, ±1.0 e up to 200,000 e and ±1.5 e above; for class II, the steps are at 5,000 e and 20,000 e. In service the limits are twice those on initial verification. [1]

Editorial calculation: a balance with Max 220 g and e = 1 mg has n = 220,000, so it falls in class I. Its minimum capacity is 100 e = 100 mg, and its initial-verification limits are ±0.5 mg up to 50 g, ±1.0 mg from 50 to 200 g and ±1.5 mg above 200 g, doubling to ±1, ±2 and ±3 mg in service. [1]

Give the supplier the capacity, readability and verification interval, accuracy class or approval required, sample sizes and required weighing tolerance, draught shield and pan size, internal or external adjustment, interfaces, and calibration certificate requirements.

02

Plan calibration: repeatability, eccentricity and minimum weight

EURAMET Calibration Guide No. 18 calls for the instrument to be warmed up, levelled and exercised before calibration, with suitable air currents, vibration and site stability. The repeatability test loads the same test load at least 5 times, for example at about half of Max; the eccentricity test places a load of about one third of Max at the centre and four off-centre positions; and errors of indication are tested at five or more loads across the range. [2]

The guide defines minimum weight as the smallest sample for which the relative measurement uncertainty equals the user's required accuracy, for example 0.1%, and says a safety factor is applied because of draughts, vibration, operators, static charge and similar influences. [2]

Editorial calculation: if the expanded uncertainty near zero load were 0.2 mg, a 0.1% requirement would give a minimum weight of 0.2 mg ÷ 0.001 = 200 mg, and a safety factor of 2 would raise it to 400 mg. The uncertainty value here is illustrative; use the figure from your balance's calibration. [2]

For pharmaceutical work, USP general chapter 〈41〉 Balances (text of the Interim Revision Announcement official 1 July 2014) applies when a substance must be "accurately weighed". Repeatability is tested by weighing one test weight at least 10 times and passes if two times the standard deviation, divided by the smallest net weight the user plans to weigh, is no more than 0.10%; a standard deviation below 0.41 d is replaced by 0.41 d, where d is the scale interval. Accuracy passes if the reading is within 0.10% of a test weight of 5–100% of capacity, whose maximum permissible error or calibration uncertainty is no more than one-third of that limit. Check the edition in force for your site. [3]

Editorial calculation: for a balance with d = 0.1 mg, the 0.41 d floor gives a standard deviation of at least 0.041 mg, so the smallest net weight that can pass is 2 × 0.041 ÷ 0.001 = 82 mg; with a measured standard deviation of 0.1 mg it becomes 200 mg. [3]

04

Analytical 0.1 mg and precision 1 mg to 0.1 g balances from listed suppliers

Stuccler and Drawell list 0.1 mg analytical and 1 mg to 0.1 g precision balances; Kuanyi lists electronic, analytical and precision balances with industrial scales; Worner, Labtech, Lachoi and Yuke list analytical or precision balances with other laboratory equipment; Biobase, Infitek and Alltrustedmed list laboratory balances within wider laboratory or medical ranges.

05

Weighing cell, pan, draught shield, levelling and adjustment

Three of these parts map straight onto calibration checks in EURAMET Calibration Guide No. 18: the pan onto the eccentricity test, the levelling feet onto set-up before calibration, and the adjustment onto the state the balance is calibrated in. [2]

PartRoleFix in the order
Weighing cellSets readability and verification interval. [1]Cell type, e and d.
Weighing panLoad position affects eccentricity error. [2]Pan size.
Draught shieldShields the pan from air currents.Shield type.
Levelling feet and indicatorInstrument must be levelled. [2]Level indicator.
AdjustmentAdjustment before use or calibration. [2]Adjustment method.

06

Capacity, d and e, repeatability, accuracy test and stabilization for a balance order

Readability alone does not set the class: OIML R 76-1 classifies by the verification interval e and n = Max/e, which A&D's legal-for-trade models state separately, for example 0.001 g read with e = 0.01 g. [1] [4]

FieldWhat it means and published valuesAsk the supplier for
Weighing capacity (g)Max; n = Max/e sets the class. [1]Max and Min.
Readability (mg)Display interval d; e may differ, d < e ≤ 10 d (e.g. d = 1 mg, e = 10 mg). [1] [4]d and e.
Repeatability (SD) (mg)At least 5 loadings (EURAMET); USP 〈41〉 NLT 10, 2 s ÷ smallest net ≤ 0.10%. [2] [3]SD and test load.
Linearity error (mg)Error of indication across ≥5 loads. [2]Errors and uncertainty.
Accuracy testUSP 〈41〉: within 0.10% with a 5–100% capacity weight. [3]Test weight and limit.
Stabilization time (s)Time to stable reading; one maker lists 1 s in fast mode. [4]Time.
Operating temperature (°C)Environment.Range.
Relative humidity (% RH)Environment.Range.
Power supply (V AC)Mains adapter.Voltage.
Power consumption (W)Mains power; allow the warm-up time. [2]Power and warm-up time.
IP ratingWashdown or dust.IP rating.
Pan size (mm)Affects eccentricity loading. [2]Diameter or size.
Net weight (kg)Balance mass.Mass.

07

Send the sample, tolerance and calibration data together

Paste it into your enquiry below or into an email to a maker, then replace every bracket.

Samples: [smallest and largest mass; type; static]
Tolerance: [required relative accuracy; minimum weight]
Balance: [Max; d; e; class or approval]
Features: [draught shield; internal adjustment; interfaces]
Calibration: [certificate; repeatability; eccentricity]
Site: [bench; vibration; air currents; temperature]

See manufacturer records Send an enquiry through CNFX

Compare precision balance suppliers

10 companies list this product among what they make. Company figures are quoted from each company's own website; every card states where the relationship came from.

Stuccler
Manufacturing site to be confirmed for the order

Lists 0.1 mg analytical, 1 mg high-precision and 0.1–0.01 g precision balances.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Chongqing Drawell Instrument Co., Ltd.
Manufacturing site to be confirmed for the order

Lists 0.1 mg analytical, 1 mg precision and 0.01 g electronic balances.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Xiamen Kuanyi Electronic Technology Co., Ltd.
Manufacturing site to be confirmed for the order

Lists 0.01 g electronic, analytical and precision balances with industrial scales.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Shaoxing Worner Lab Equipment Co., Ltd.
Manufacturing site to be confirmed for the order

Lists analytical and precision balances with laboratory equipment.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Shanghai Labtech Co.,Ltd
Manufacturing site to be confirmed for the order

Lists precision and analytical balances with laboratory analyzers.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Lachoi Scientific Instrument (Shaoxing) Co., Ltd.
Manufacturing site to be confirmed for the order

Lists analytical balances with stirrers and water baths.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Shanghai Yuke Industrial Co., Ltd.
Manufacturing site to be confirmed for the order

Lists electronic analytical balances with laboratory instruments.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
BIOBASE GROUP
Manufacturing site to be confirmed for the order

Lists laboratory balances within a wide laboratory equipment range.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Infitek
Manufacturing site to be confirmed for the order

Lists laboratory balances within a wide laboratory equipment range.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
Alltrustedmed
Manufacturing site to be confirmed for the order

Lists laboratory balances among medical and laboratory equipment.

Product scope checked 2026-10-11

Current product source ↗

Balances listed · profile compiled by CNFX from public sources
View All Factories

Explore company information and manufacturing capabilities, then discuss the scope of your project with a relevant supplier.

Share this page

Buyer questions

What is the difference between readability and verification interval?

Readability is the display step; the verification scale interval e is used for accuracy classes and permissible errors, and may be chosen by the manufacturer within OIML limits. [1]

How large are permissible errors for a class I balance?

±0.5 e up to 50,000 e, ±1.0 e up to 200,000 e, ±1.5 e above, on initial verification; doubled in service. [1]

What is minimum weight?

The smallest sample for which relative measurement uncertainty equals the required accuracy, with a safety factor applied. [2]

What does USP 〈41〉 require of a balance?

Repeatability of 2 × SD ÷ smallest net weight no more than 0.10% over at least 10 weighings, with SD at least 0.41 d, and accuracy within 0.10% using a test weight of 5–100% of capacity. [3]

Why does a legal-for-trade balance show both d and e?

OIML R 76-1 allows e to differ from the display interval d, within d < e ≤ 10 d; A&D's GX-203AN, for example, reads to 0.001 g with e = 0.01 g. [1] [4]

Sources and what still needs a direct check

Sources checked 2026-10-12. OIML is the intergovernmental organisation for legal metrology; EURAMET is the association of European national metrology institutes; USP 〈41〉 is the US Pharmacopeia chapter on balances. None of them recommends a product. A&D's Apollo GX-A/GF-A product page is cited as an example of how a maker states d, e and built-in checks, not as an endorsement.

  1. OIML R 76-1:2006 — Non-automatic weighing instruments, Part 1: Metrological and technical requirements ↗ International recommendation Return to text ↑
  2. EURAMET Calibration Guide No. 18 v4.0 — Guidelines on the calibration of non-automatic weighing instruments ↗ Metrology institute calibration guide Return to text ↑
  3. US Pharmacopeia — General chapter 〈41〉 Balances, Interim Revision Announcement official 1 July 2014 (copy hosted by Precisa) ↗ Pharmacopeia chapter Return to text ↑
  4. A&D Weighing — Apollo GX-A/GF-A Series precision balances with NTEP models (product page: features and model list) ↗ Manufacturer technical page Return to text ↑

Still to check yourself

  • Confirm Max, d, e and class.
  • Ask for repeatability and minimum weight.
  • Plan the bench and draught protection.
  • Request a calibration certificate.
  • Verify the legal supplier and factory.
Buyer enquiry

Request manufacturing insight for Precision Balance

Ask for use case, specification boundaries, supplier type, and RFQ preparation information for this product.

Where it goes
Straight to the CNFX editorial desk, and to the manufacturer if this product is linked to a claimed profile. Nothing is broadcast to a supplier list.
Your details stay here
We do not sell or rent enquiry data, and we do not add you to a mailing list. Used only to answer this request.
No commission, no middleman
CNFX is a directory. We take no cut of any order and never negotiate on a supplier's behalf.
What to include
Share your product specifications, quantity, destination and target delivery date.

Your business information is used only to process this request.

Thank you! Your message has been sent. We'll respond within 1–3 business days.
Sorry, we couldn't send your message. Please try again, or email us at [email protected].

Need to Manufacture Precision Balance?

Compare manufacturer profiles with relevant product and process capability.