Process and laboratory calibrators · Multifunction electrical and loop calibrators, dry-block temperature wells, deadweight pressure testers, accuracy as % of reading plus floor, stability and uniformity, traceability and accredited certificates
Calibration equipment is the set of reference instruments used to check and adjust process and test instruments against a known value: field and bench calibrators that source and measure electrical signals, temperature sources such as dry-block wells with reference thermometers, and pressure standards such as deadweight testers and pressure modules. Ask three things first: which instruments and ranges must be calibrated, what accuracy or uncertainty the reference must have relative to the device under test, and what certificate and traceability chain the reference itself must carry.
Reference instruments for calibrating process and test instruments: multifunction and loop calibrators, dry-block temperature calibrators with reference probes, deadweight testers and pressure calibrators, calibration kits of such references, and automated calibration benches built from them. Calibration services, coordinate measuring machines, and the instruments being calibrated are outside this page except where compared.
01
Fluke describes the 726 as a field calibrator that measures volts, mA, pressure, RTDs, thermocouples, frequency and resistance, and sources or simulates volts, mA, thermocouples, RTDs and frequency to calibrate transmitters, on two separate channels. Its published DC voltage measurement accuracy is 0.01% + 2 mV on the 30 V range and 0.01% + 10 μV on the 90 mV range; it sources 24 mA at 0.01% + 2 μA, measures 0 to 4,000 Ω at ±0.015%, provides a 24 V loop supply limited to 22 mA, works from −10 to +55 °C, weighs 650 g and runs on four AA cells in a compartment that opens without breaking the calibration seal. Pressure is measured by attaching one of Fluke's 750P modules, from 0.025% of range. [1]
Editorial calculation: an accuracy of 0.01% of reading + 2 mV applied to a 10 V reading gives 0.001 V + 0.002 V = ±3 mV, so at mid-range the fixed floor is twice the proportional term; compare the total, not the percentage alone, with the tolerance of the instrument under test. [1]
For temperature, Fluke's 9142 field metrology well is a portable dry-block source covering −25 to 150 °C with ±0.2 °C display accuracy, ±0.01 °C stability, ±0.01 °C radial (well-to-well) uniformity, ±0.05 °C axial uniformity at 40 mm, 150 mm immersion depth, about 25 minutes from −25 to 150 °C, and a weight of 8.16 kg. Fluke lists accuracy, stability, axial and radial uniformity, loading and hysteresis as the six criteria for such wells, states each unit ships with an accredited calibration certificate and a 4:1 test uncertainty ratio, and offers a process option with a built-in reference-thermometer readout and 4–20 mA input with 24 V loop power. [2]
Give the supplier the instruments to be calibrated (signal types, ranges, sensor types), the tolerance of each, the required test uncertainty ratio, field or laboratory use, temperature span and probe diameters and lengths, need for loop power and HART, documentation and data transfer, and the calibration certificate type required with delivery.
02
Fluke describes deadweight testers as piston-gauge pressure standards: calibrated masses are loaded on a precision-lapped piston that floats in its cylinder, and the pressure is the sum of the weights and the carrier divided by the piston area. Its P3100 hydraulic series reaches 20,000 psi (1,400 bar) in single or dual piston form, with ±0.015% of reading accuracy (±0.008% optional) from 10% to 100% of the piston range; below 10% the error is the accuracy class times 10% of the piston range. Pistons are tungsten carbide, standard masses non-magnetic austenitic stainless steel, and each unit can be calibrated to local gravity, otherwise to standard gravity of 980.665 cm/s². [3]
Editorial calculation: on a 1 to 35 bar piston at the 0.015% class, the error below 10% of range (3.5 bar) is held at 0.015% × 3.5 bar ≈ 0.53 mbar, so a 1 bar point carries about 0.05% of reading; choose a piston range that keeps the test points above 10% of it. [3]
NIST adopts the VIM definition of metrological traceability as a measurement result related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty. It states that supporting a traceability claim is the responsibility of whoever provides the result, that assessing it is the user's responsibility, that NIST does not certify the traceability of results other than its own, and that traceability alone does not show fitness for purpose because the uncertainty must also be small enough for the measurement need. [4]
Ask for the calibration certificate of each reference: the laboratory and its accreditation, the values and uncertainties reported, the reference standards used, the gravity value for deadweight testers, and the recommended recalibration interval.
03
| Equipment | Checks | Key figures to compare |
|---|---|---|
| Multifunction process calibrator | Transmitters, loops, RTD and thermocouple inputs, switches. [1] | Accuracy per range (% + floor), loop power, channels. |
| Dry-block temperature well | Temperature sensors and transmitters by comparison. [2] | Range, stability, uniformity, immersion depth, reference readout. |
| Deadweight tester | Pressure gauges and transmitters. [3] | Range, accuracy class, gravity correction, medium. |
| Pressure module | Pressure with an electrical calibrator. [1] | Range type (gauge, differential, absolute), % of range. |
04
| Part | Role | Fix in the order |
|---|---|---|
| Reference standard | Carries the traceable value: PRT, piston and masses, internal references. [2] [3] | Certificate and uncertainty. |
| Source or simulator | Generates the signal, temperature or pressure. [1] [2] [3] | Range, stability, resolution. |
| Readout | Measures the device under test. [1] [2] | Accuracy, channels, loop power. |
| Interface and software | Records as-found and as-left data. [2] [3] | RS-232, software, report format. |
| Certificate | Documents traceability. [4] [2] | Accreditation, uncertainty statement. |
05
Figures come from the Fluke 726, 9142 and P3100 pages; ingress protection and a recalibration interval were not published there, so both have to come from the supplier.
| Field | What it means and published values | Ask the supplier for |
|---|---|---|
| Accuracy | 0.01% + 2 mV (30 V); ±0.2 °C display; ±0.015% of reading. [1] [2] [3] | Accuracy per range with conditions. |
| Measurement range | Up to 30 V DC, 24 mA; −25 to 150 °C; to 1,400 bar. [1] [2] [3] | Ranges you need. |
| Resolution | 0.01 °C RTD; 0.01° display. [1] [2] | Resolution per range. |
| Stability | ±0.01 °C full range in one well. [2] | Short-term and annual stability. |
| Operating temperature | −10 to +55 °C; 0 to 50 °C. [1] [2] | Range over which specs hold. |
| Relative humidity | 0 to 90% RH non-condensing. [2] | Humidity limits. |
| Power supply | Four AA cells; 100–115 V or 230 V AC. [1] [2] | Supply and power. |
| Ingress protection | Not stated in the reviewed sources. | IP rating if used outdoors. |
| Weight | 650 g; 8.16 kg; 16 kg. [1] [2] [3] | Weight with case. |
| Dimensions (W×D×H) | 200 × 96 × 47 mm calibrator; 290 × 185 × 295 mm well (H × W × D). [1] [2] | Size and case. |
| Calibration cycle | Not stated in the reviewed sources. | Recommended interval. |
| Warm-up time | Heating 25 min from −25 to 150 °C in one well. [2] | Warm-up and settling time. |
06
Paste it into your enquiry below or into an email to a maker, then replace every bracket.
Instruments: [types; ranges; sensor types; tolerances] Ratio: [required test uncertainty ratio] Temperature: [span; probe diameter and length] Pressure: [range; gauge or absolute; medium] Use: [field or laboratory; documentation] Certificate: [accredited; uncertainty; gravity value] Supply: [quantity; cases; service]
Xi'an Shelok Instrument Co.,Ltd lists dead weight testers, pressure calibrators and dry-block temperature calibrators; JIANGSU HONGGUANG INSTRUMENT FACTORY CO,LTD lists a loop calibrator.
Lists dead weight testers and pressure calibrators, with dry-block temperature calibrators and thermostatic baths in its temperature calibration range.
Product scope checked 2026-10-13
Lists a loop calibrator that measures and sources voltage and current signals and supplies transmitter loop power.
Product scope checked 2026-10-13
Explore company information and manufacturing capabilities, then discuss the scope of your project with a relevant supplier.
A proportional term on the reading plus a fixed floor; at 10 V the total is about ±3 mV. [1]
Fluke names stability, axial and radial uniformity, loading and hysteresis alongside accuracy. [2]
The force of the masses depends on local gravity; Fluke calibrates to local gravity on request, otherwise to standard gravity. [3]
No. NIST notes traceability alone does not show fitness for purpose; the uncertainty must also suit the measurement. [4]
Sources checked 2026-10-11. The Fluke pages describe one maker's calibrators, temperature wells and deadweight testers; NIST's policy describes what traceability means. Values for your order must come from the supplier offered and its certificates.