Reference-class bridge CMMs · sub-micron E0, form measurement, 20–22 °C rooms, Zerodur scales, probe and stylus changing
A high-accuracy reference CMM is a bridge coordinate measuring machine built and specified for sub-micron work: calibrating gauges and test specimens, measuring roundness and cylindricity, and serving as the reference in a measuring laboratory. Ask three things first: the ISO 10360 errors the maker will accept at your lengths, the temperature band and gradients that hold them, and the probing and stylus changing the machine will use.
Bridge CMMs specified in the sub-micron class for calibration, form measurement and reference measuring rooms, with their scanning probes and stylus or module change racks. Production bridge, gantry and similar Cartesian CMMs with general ISO 10360-2 acceptance are covered on the Coordinate measuring machines (CMM) page; portable arms, laser trackers and vision machines are outside this page.
01
ZEISS states a length measurement error of 0.5 + L/500 µm for PRISMO ultra and positions it for research, quality assurance and the calibration of gauges or test specimens in reference measuring rooms; the machine uses ZERODUR scales with 20 nm resolution, thermal bending compensation, active pneumatic damping and improved air bearings. [1]
TNO publishes the figures for its own installed PRISMO ultra: E0 = 0.5 + L/500 µm at 20–22 °C and repeatability range R0 of 0.4 µm under ISO 10360-2:2009, scanning error THP of 0.9 µm under ISO 10360-4:2000, single-stylus probing error PF of 0.5 µm and multi-stylus errors PFTH 1.9 µm, PSTM 0.6 µm and PLTM 1.2 µm under ISO 10360-5:2010, a roundness form measurement error of 0.5 µm and a probing form tolerance PFTU of 1.0 µm. [2]
Editorial calculation: with E0 = 0.5 + L/500 µm, the permitted length error is 1.5 µm at L = 500 mm and 2.5 µm at L = 1000 mm, so the constant term dominates only on short lengths. [1] Ask for every figure in the set, not E0 alone, because multi-stylus and scanning errors on TNO's machine are larger than its single-stylus PF. [2]
ZEISS states that with its optional form measurement package, which includes a special probe set, the machine guarantees calibration of a ring gauge to 0.2 µm or better, and that roundness and cylindricity tolerances from 2 to 3 µm can be measured under the golden rule of measurement technology. [1]
Give the supplier the features and tolerances you must measure, the longest lengths, the probing method (single point, scanning, multi-stylus), and the ISO 10360 parts and values you will use for acceptance.
02
ZEISS lists 20–22 °C for PRISMO ultra, and TNO quotes its E0 at 20–22 °C. [1] [2] TNO's machine stands in a cleanroom classed 6/7 with an ESD-safe floor. [2]
ZEISS states that every PRISMO ultra granite plate is acclimatized for three months before delivery, the final inspection takes place in an air-conditioned chamber, and after installation a precision test is followed 4 to 6 weeks later by an acceptance test by an expert. [1] Editorial note: build that interval into the commissioning plan; the machine is not accepted on the day it is installed.
TNO's machine has a measuring volume of 900 × 1300 × 650 mm, a table height of 860 mm, a machine weight of 2770 kg and a maximum workpiece mass of 1500 kg; ZEISS lists sizes from 700 × 1000 × 500 mm to 1600 × 3000 × 1000 mm. [2] [1] Editorial calculation: a 2770 kg machine carrying a 1500 kg part puts 4270 kg on the floor, which the foundation and any isolation must carry.
03
Renishaw's SP25M combines a scanning and a touch-trigger sensor in a 25 mm diameter housing, with five scanning modules carrying M3 styli from 20 to 400 mm and an adaptor for TP20 touch-trigger modules. [3]
Renishaw's FCR25 change racks store any SP25M element in each port, as a triple-port unit for its MRS modular rack or as 3-port and 6-port stand-alone units. [3] TNO's PRISMO ultra carries the ZEISS VAST Gold active scanning probe head. [2] Ask which probe and rack the quoted accuracy figures were measured with.
04
| Part | Role | Fix in the order |
|---|---|---|
| Granite table | Base of the measuring volume; acclimatized three months before delivery on PRISMO ultra. [1] | Size, load and flatness certificate. |
| Bridge and air bearings | Carry the axes; improved air bearings on PRISMO ultra. [1] | Bearing type and air requirement. |
| Scales | ZERODUR scales, 20 nm resolution. [1] [2] | Scale material and resolution. |
| Damping | Active pneumatic damping. [1] | Isolation for your floor. |
| Probe head and probe | Active scanning head; SP25M scanning and touch-trigger modules. [2] [3] | Probe used for the accuracy figures. |
| Change rack | FCR25 3- or 6-port units. [3] | Number of ports and adaptors. |
| Rotary table (option) | ZEISS cites a VAST rotary-table axis option. [1] | Accuracy and load. |
05
| Field | What it means and published values | Ask the supplier for |
|---|---|---|
| Measurement Volume (X×Y×Z) (mm) | 700 × 1000 × 500 to 1600 × 3000 × 1000 mm (ZEISS); 900 × 1300 × 650 mm at TNO. [1] [2] | Volume for your largest part. |
| Maximum Probing Error (μm) | PF 0.5 µm single stylus; PFTH 1.9 µm multi-stylus (TNO). [2] | ISO 10360-5 values with your probe. |
| Scanning Speed (mm/s) | ZEISS doubled PRISMO-family scanning speed from 150 to 300 mm/s. [1] | Speed at which THP is stated. |
| Positioning Accuracy (μm/m) | Length error E0 = 0.5 + L/500 µm. [1] | E0 and R0 values for acceptance. |
| Maximum Workpiece Weight (kg) | 1500 kg on TNO's machine. [2] | Load and its distribution. |
| Environmental Temperature Range (°C) | 20–22 °C. [1] [2] | Band and gradient limits. |
| Resolution (mm) | 20 nm scale resolution. [1] | Scale and system resolution. |
| Repeatability (mm) | R0 0.4 µm (TNO). [2] | R0 under ISO 10360-2. |
| Maximum Acceleration (mm/s²) | Axis dynamics. | Value. |
| Power Supply (V AC) | Mains for controller. | Voltage and phase. |
| Air Supply Pressure (MPa) | For air bearings and damping. | Pressure, flow and air quality. |
| Machine Weight (kg) | 2770 kg at TNO. [2] | Weight and foundation drawing. |
| Footprint (L×W×H) (mm) | Table height 860 mm at TNO. [2] | Footprint and clearances. |
| Ingress Protection Rating | Lab machine; TNO's sits in a class 6/7 cleanroom. [2] | Room requirements. |
06
Paste it into your enquiry below or into an email to a maker, then replace every bracket.
Parts: [largest size; weight; material] Tolerances: [lengths; form tolerances; ratio required] Accuracy: [E0; R0; PF; THP; multi-stylus values] Probing: [scanning or touch; probe head; stylus lengths; change rack ports] Room: [temperature band; gradients; floor; cleanliness] Acceptance: [ISO 10360 parts; who tests; when]
Manufacturer profiles associated with Precision Industrial Coordinate Measuring Machine.
Explore company information and manufacturing capabilities, then discuss the scope of your project with a relevant supplier.
ZEISS states E0 = 0.5 + L/500 µm for PRISMO ultra; at 1000 mm that is 2.5 µm by editorial calculation. [1]
TNO lists R0 0.4 µm, THP 0.9 µm, PF 0.5 µm, PFTH 1.9 µm, PSTM 0.6 µm and PLTM 1.2 µm for its machine. [2]
Renishaw's SP25M modules carry M3 styli from 20 to 400 mm across five modules. [3]
Sources checked 2026-10-12. ZEISS and Renishaw describe their own machine and probe; TNO, the Netherlands research organisation, publishes the specification of the machine in its laboratory.