Process weighing · Load cell capacity and count, OIML accuracy classes, wiring, piping and calibration
A load cell scale turns a tank, hopper, vessel or platform into a weighing system by supporting it on load cells or weigh modules wired to a junction box and indicator. Ask three things first: the dead load, live load and number of supports that set load cell capacity, the accuracy class and resolution you need, and how piping, structure and cables will be arranged so they do not carry weight.
Load cells, weigh modules, junction boxes and indicators assembled into tank, hopper, vessel and process platform scales. Truck and axle scales, retail and personal scales, laboratory balances and calipers are outside this page.
01
Mettler Toledo's weigh module guidance says load cells measure both dead load and net load, and that using load cells larger than needed is not recommended because accuracy suffers. For scales where the goods sit centred over the supports, such as tanks, silos and vessels, it gives C ≥ sf × (TDL + SC) ÷ N; for four-cell floor, small-vehicle and conveyor scales with off-centre loads, C ≥ sf × (TDL ÷ 4 + SC ÷ 2), where C is load cell or weigh module capacity, TDL total dead load, SC scale capacity, N the number of cells and sf a safety factor, which Mettler Toledo gives as 1.25. [1]
Editorial calculation: a tank with 800 kg dead load and 2,000 kg scale capacity on 3 weigh modules needs C ≥ 1.25 × (800 + 2,000) ÷ 3 ≈ 1,170 kg per module; on 4 modules, about 875 kg. Choose the nearest standard capacity at or above the result rather than a much larger cell. [1] The dead and live loads here are illustrative.
Mettler Toledo describes tank, hopper and vessel applications as static loading with little horizontal shear on the cells, while conveyors, pipe racks and high-powered mixers impose dynamic loads that need suitable modules. [1]
Give the supplier the vessel drawing, empty and full weights, number and position of supports, product and process (agitation, heating, pressure), required resolution and accuracy, piping and connections, environment (washdown, hazardous area, outdoor wind or seismic), indicator outputs and calibration method.
02
OIML R 60 ranks load cells in classes A, B, C and D by the maximum number of verification intervals: class C covers 500 to 10,000 intervals, for example C3 for 3,000; class B 5,000 to 100,000; class D 100 to 1,000. Its maximum permissible errors are an apportioning factor (0.7 unless declared) times 0.5, 1.0 or 1.5 verification intervals depending on load. The Recommendation notes that a load cell need not be of the same class as the measuring system it is used in. [2]
On wiring, Mettler Toledo says each cell's cable runs to a junction box that sums the signals for the indicator, an analog junction box sums up to four cells, cables should not be shortened or lengthened in the field because that changes the output, and 6-wire cells add sense lines that help where the indicator is far from the scale. Its handbook treats tank piping-connection design together with the support structure and foundation. [1]
Pipework is where a well-chosen cell loses accuracy. At the European Commission's Joint Research Centre in Ispra, a 400 dm³ D-shaped product tank on a weighing system for nuclear material accountancy was chosen for an accuracy potential of 1 in 10⁵ and to avoid density, dip-tube and temperature problems of volume methods. The team commissioned it first with no pipes attached, then measured the influence of pipe connections and of off-centre loading as the centre of gravity moved during filling; after modifying the pipework, optimising its clamping and recalibrating, initial results with process pipework attached showed a repeatability of ±0.006% of 400 kg, the liquid weight in the tank. [4]
Editorial calculation: ±0.006% of 400 kg is ±24 g; for comparison, a 3,000-interval scale (OIML class C3 cells) on a 400 kg span has an interval of about 133 g, so the JRC result after pipe clamping and recalibration sits well inside one such interval. [4] [2]
03
Flintec's US patent 6,172,309 (filed 6 April 2000, granted 9 January 2001) describes the shear beam cell used in weigh modules: a rectangular beam of load-cell-quality aluminium, steel or stainless steel alloy, bolted at one end as a cantilever and loaded through a vertical hole at the free end. Between the two, opposed lateral recesses leave an I-section with a thin vertical web; strain gauges bonded to both sides of the web measure principal strain due to shear and are connected in a bridge circuit. [3]
The patent explains that the vertical shear force is the same in every cross section between the support edge and the loading hole, so shear stress varies inversely with the cross section; capacity ratings in a range are set by changing the web thickness while the blank and recess sizes stay constant. Known shear beam designs, it says, give high accuracy with low sensitivity to off-centre loads, torque and horizontal forces at the load point. [3]
Its design cuts the web pockets with a side-cutting tool so the web thickens gradually away from a flat centre where the gauges sit, giving more uniform shear strain there and lower off-centre load sensitivity, and reduces machining from five set-ups for the prior design to one. [3] Editorial note: for a weigh module, ask the supplier for the cell type (shear beam, bending beam, column), body material and sealing, the OIML R 60 class and the off-centre load test results, alongside the capacity worked out with Mettler Toledo's formula. [2] [1]
04
Suppliers linked to this page list load cells, weigh modules or indicators for weighing systems; each card links to the page reviewed.
Zhihe lists pancake, shear beam and tension load cells with weigh modules, junction boxes and indicators; Hongli lists OIML/NTEP shear beam load cells and weighing modules with junction boxes and indicators; Top Sensor lists load cells, weight indicators and load cell amplifiers; Kuanyi lists load cells, weighing indicators and junction boxes with scales.
05
This is a functional reference built from the sources cited; it is not one maker's parts list.
| Part | Role | Fix in the order |
|---|---|---|
| Load cell or weigh module | Measures dead and net load; class per OIML R 60. [1] [2] | Capacity, class, type. |
| Mounting hardware | Introduces load; may restrain tipping. [1] | Module type and restraints. |
| Cables | Fixed length; do not cut or extend. [1] | 4- or 6-wire; length. |
| Junction box | Sums up to four analog cells. [1] | Box size and sealing. |
| Indicator | Reads the summed signal as weight. [1] | Outputs and approvals. |
06
A weighing system quotation needs values for your vessel and process. The table explains each field, gives published values where a reviewed source states them, and says what to obtain from the supplier.
| Field | What it means and published values | Ask the supplier for |
|---|---|---|
| Capacity (kg) | Per cell ≥ 1.25 × (dead + live) ÷ N for centred loads. [1] | Cell capacity and count. |
| Accuracy (g) | System accuracy from class and intervals. [2] | Class and system accuracy. |
| Response time (ms) | Indicator filtering. | Update rate. |
| Operating temperature (°C) | Class marking may state the range, e.g. +5 to +35 °C. [2] | Range. |
| Output signal | Analog summed in junction box, or digital. [1] | Signal and protocol. |
| Rated output (mV/V) | Sensitivity. | Value and tolerance. |
| Excitation voltage (V DC) | Supply from indicator. | Voltage. |
| Input resistance (Ω) | Bridge. | Value. |
| Output resistance (Ω) | Bridge. | Value. |
| Insulation resistance (MΩ) | Bridge to body. | Value. |
| Safe overload (% FS) | Above rated capacity. | Value. |
| Ultimate overload (% FS) | Damage limit. | Value. |
| Ingress protection | Washdown. | IP rating. |
| Material | Load cell body; aluminium, steel or stainless steel alloy named in one patent. [3] | Material and finish. |
| Piping and commissioning | Test with no pipes, then with pipes; clamp and recalibrate (JRC Ispra: ±0.006% of 400 kg repeatability after pipework optimisation). [4] | Commissioning test plan with pipework connected. |
07
Paste it into your enquiry below or into an email to a maker, then replace every bracket.
Vessel: [drawing; supports; empty weight; full weight] Process: [static or dynamic; agitator; temperature; pressure] Accuracy: [resolution; class; legal-for-trade or not] Installation: [piping connections; support structure; wind/seismic] Electronics: [junction box; indicator; outputs; cable lengths] Calibration: [test weights; method; certificate]
These companies list load cells, weigh modules or indicators.
Lists load cells, weigh modules, junction boxes and weighing indicators.
Product scope checked 2026-10-11
Lists OIML/NTEP load cells, weighing modules, junction boxes and indicators.
Product scope checked 2026-10-11
Lists load cells, weight indicators and load cell amplifiers.
Product scope checked 2026-10-11
Lists load cells, weighing indicators and junction boxes with scales.
Product scope checked 2026-10-11
Explore company information and manufacturing capabilities, then discuss the scope of your project with a relevant supplier.
Mettler Toledo gives capacity per cell ≥ 1.25 × (dead load + scale capacity) ÷ number of cells for centred loads, and warns against oversizing. [1]
OIML R 60 class C with 3,000 verification intervals. [2]
Mettler Toledo says not to lengthen or shorten load cell cables in the field because it changes the output; coil excess cable near the junction box. [1]
Flintec's US patent 6,172,309 describes strain gauges bonded to both sides of a thin web in a cantilevered beam, measuring shear strain in a bridge circuit; capacity is set by web thickness. [3]
Yes. JRC Ispra commissioned a 400 dm³ weighing tank without pipes, then measured the influence of pipe connections; after re-clamping the pipework and recalibrating, it reached a repeatability of ±0.006% of 400 kg. [4]
Sources checked 2026-10-11. OIML is the intergovernmental organisation for legal metrology. Mettler Toledo and Flintec are weighing equipment makers; Mettler Toledo's weigh module guidance and Flintec's load cell patent describe their own practice. The JRC Ispra paper reports commissioning tests on one weighing tank.